FDA MedWatch - BagEasy Manual Resuscitation Devices by Westmed, Inc.: Class 1 Recall

Westmed, Inc. is initiating a nationwide recall of 24,384 units of BagEasy Manual Resuscitation Devices.

AUDIENCE: Risk Manager, Emergency Medicine

ISSUE: Westmed, Inc. is initiating a nationwide recall of 24,384 units of BagEasy Manual Resuscitation Devices. The BagEasy device have been found to have a potential for disconnection at the retention ring of the patient port manifold. Disconnection causes the unit to be inoperable, which potentially could result in treatment delays while another unit is obtained or technician switches to a different method of resuscitation.

BACKGROUND: Westmed has become aware of a potential for disconnection at the patient port retention ring assembly of the BagEasy manual resuscitation device after receiving reports from three separate facilities regarding units disconnecting during setup or use. The BagEasy manual resuscitation device was distributed to other medical device distribution companies as well as directly to hospitals and other treatment facilities.

RECOMMENDATIONS: Consumers with questions may contact the company at 1-800-975-7987 and arrange for return of all recalled products. For the affected part numbers and lot numbers please see the firm press release. Healthcare professionals are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program:

* Complete and submit the report Online: www.fda.gov/MedWatch/report.htm
* Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178.

Liked this article? Start your subscription to Healthcare Packaging for FREE:

Sign up to receive the print magazine six times per year and the e-mail newsletter twice a month.

First Name:
Last Name:
 
Company:
Email Address:

 

* indicates a sponsored article that was submitted directly to this Web site by the supplier, and was not handled by the Healthcare Packaging editorial staff. Healthcare Packaging may share your contact information with our sponsors, as detailed in our Privacy Policy. Healthcare Packaging will not share your information with a sponsor whose content you have not reviewed.







Newsletters

Don’t miss intelligence crucial to your job and business!
Click on any newsletter to view a sample. Enter your email address below to sign up!
HEALTHCARE
PACKAGING MACHINERY

On the Edge

Keith Campbell speaks out!

New Machines

New packaging machines

FROM SISTER PUBLICATION AUTOMATION WORLD

Process Automation

Industry Trends & Applications

GENERAL INTEREST

eClip

Breaking packaging news

Packaging Insights

Pertinent packaging issues

PACKAGE DESIGN/
DEVELOPMENT

Greener Package

Sustainable packaging

Shelf Impact

Package design strategies

PACK EXPO
Each newsletter ranges in frequency from once per month to a few times per month at most.
Email: